Random access, discrete automatic biochemistry analyzer, insert emergency at any time
High performance and reasonable price, It is the best economical choice from semi-auto upgraded to fully-auto analyzer
Probe performance: Liquid level detection, all-round anti-collision protection and inventory checking, have automatic stop and alarm prompt for an obstacle
Probe cleaning: Automatic washing for both interior and exterio
Analysis System
Test Speed: constant 160tests/hour, up to 220tests/hour
Test method: End-point, Fixed-time (two point method), kinetic
Reaction tray: 81 UV cuvettes (disposable)
Mini reaction volume: 150μl, save reagent cost
Insert emergency at any time
Support linear and non-linear calibration
Heating method: solid constant temperature closed direct heating
Temperature equilibrium time: less than 3minutes,
Reaction Temperature: 37±0.2°C
Temperature Fluctuation: ± 0.2°C
Optical System
Integrated full closed optical system, after spectrometric method, no mechanical parts, not moldy, no maintenance cost
8 wavelengths, selectable among 340-800nm
Absorption range: 0.00-4.00Abs
Halogen-tungsten lamp has more than 2000hrs service time
Center wavelength deviation: ±2nm, half wave width less than 12nm
Inspection and Judgment: Reaction end point judgment, reaction linear range judgment, substrate depletion judgment, reaction limit inspection, reagent blank absorbance judgment
RS232 standard interface, support the intranet and remote printing
Technical Specifications
Performance
Type: Discrete Automatic, Random Access
Principle: Absorption Spectrometry, Turbidimetry
Throughput: Constant 160tests/hour (without ISE)
Test method: End-point, Fixed-time, Kinetic
Support dual reagents/double wavelength
Test sequence: Run tests in strict accordance with sample sequence,
Insert emergency at any time
Optical System
Light Source: Halogen-tungsten lamp, 12V/20W
More than 2000hrs service time
Spectrometric Method: After spectrometric method
Wavelength: 8 wavelengths, selectable among 340-800nm
Absorption Range: 0-4.0Abs
Optical System: Integrated full closed optical system
Reaction system
Reaction Tray: 81 positions for cuvettes
Reaction Volume: Mini 150μl
Reagent Tray: 60 positions, open system
Reagent Volume: 20-500μl, 1μl by step
Reagent Bottle: Bottles with 15ml,20ml,30ml,50ml,70ml are available dead volume<1ml
Temperature Control System
Heating Mode: Solid constant temperature closed direct heating
Equilibrium Time: less than 3minutes
Reaction Temperature: 37±0.2°C
Temperature Fluctuation: ± 0.2°C
Refrigeration system: For option
Calibration and QC
Calibration Mode: Linear, Logit-4p, Logit-5p, Spline, Factor, Polynomial
Automatically check calibration curve and select the calibration type with the best fitting
QC Type: Real time QC, QC in one day, QC between days;
QC Rules: QC chart: L-J, Cumulative, Twin Plot, Multi-rule; default: Westguard
Status Monitoring: Real -time monitor the sample tray, reagent tray, reaction tray, reaction cup, lamp, purified water; Real-time display reaction tray humidity, reagent remain, reaction curve, calibration curve and QC chart.
Inspection & Judgment: Reaction end point judgment, reaction linear range judgment, substrate depletion judgment, front belt detection,reaction limit inspection, reagent blank absorbance judgment
Operation and Print
Operation System: Windows XP or Windows 7 Home
Database: SQL Server
Data Storage: Automatic storage and backup, permanently preserve all kinds of data
Print Content: Print content can be customized
Print Format: Provide a variety of print format
Working Conditions
Water Consumption: less than 3L/hour
Power Supply: AC 220 ±22V 50HZ ±1HZ
Power: Max 200VA
Ambient Temperature: 10°C-30°C
Dimension(L*W*H): 78*58*55cm
Net Weight: 45K